Zahlavi

HK4 - Strain scanner

The two-axis diffractometer SPN-100 is a 2-axis neutron diffractometer usually used for mapping of residual strains inside polycrystalline materials.Typical examples are residual strains developed in the course of welding or strains accumulated during processing and/or usage of engineering components.

Instrument description

SPN-100 is a double axis diffractometer employing elastically bent perfect Si crystal as a focusing monochromator. The instrument is designed for measurements of residual strains by accurate evaluation of positions of a selected diffraction peak from small material volume defined by a set of input and output slits. The diffractometer has a changeable monochromator take off angle and can be set and operated at a suitably chosen wavelength in the range 0.1 nm to 0.235 nm. In the case of α-Fe and γ-Fe samples, it usually  operates at the neutron wavelength of 0.23 nm, providing a good resolution after diffraction on the α-Fe(110) and / or γ-Fe(111) lattices planes.

Layout

HK4 InstrumentHK4 Layout

Parameters

Type of instrument Two-axis diffractometer
Source LVR-15, 10 MW research reactor
Beam tube Horizontal channel No. 4
Monochromators Bent Si(111) and Si(220)
Take-off angle (range) 30 – 60°
Neutron flux on specimen 2 × 105 cm−2·s−1
Incident wavelength 1.0 – 2.35 Å
Detector 2-D Position Sensitive Detector, 3He, active area of 230 × 230 mm2, resolution 2 x 2 mm2.
Sample - detector distance 1000 mm
Beam defining optics Cd slits in the incident and diffracted beams
Specimen weight capacity XYZ stage: up to 50 Kg
Robotic arm: up to 7 Kg

Sample positioning

For sample positioning, in addition to standard x-y-z translation stage, a six-axis robotic arm is available allowing more flexible manipulation of complex-shaped samples and automation of strain measurements.

Sample positioning by robotic arm video

Selected publications

  • Correlation of magnetic properties and residual stress distribution monitored by X-ray and neutron diffraction in welded AISI 1008 steel sheets. Vourna, P., Hervoches, C., Vrana, M., Ktena, A., Hristoforou, E. (2015) IEEE Transactions on Magnetics, 51 (1), art. no. 7029219.
  • Residual stresses determination by neutron diffraction in a 100Cr6 chromium steel bearing ring. Rogante, M., Martinat, G., Mikula, P., Vrána, M. (2013) Kovove Materialy, 51 (5), pp. 275-281.
  • Residual stresses assessment in coated materials: Complementarity between Neutron and X-ray techniques. Rogante, M., Mikula, P., Vrána, M. (2011) Key Engineering Materials, 465, pp. 259-262.
  • Suppression of surface effect by using bent-perfect-crystal monochromator in residual strain scanning. Vrána, M., Mikula, P. (2005) Materials Science Forum, 490-491, pp. 234-238.
  • High-resolution neutron diffraction for nondestructive analysis of residual stresses in polycrystalline materials. Mikula, P., Lukáš, P., Vrána, M. (2005) Applied Mechanics and Materials, 3-4, pp. 331-336.

Applications for beamtime

All experiment proposals are evaluated on a continuous basis, they have to be submitted via the CANAM user portal.
Potential users are encouraged to contact instrument responsible before submission.